Tip of the Week
January 10, 2005
Click HERE for Adobe version of this Tip
Things about Method
17 that you should know…
…but were afraid to ask
- Always
inspect the top of the thimble
before use.
- Prior
to assembling the sampling train, pre-wet
the inner surface of the gum rubber sampling line to minimize subsequent
absorption of sample moisture.
- After
the pre-test leak check, verify that you can pull gas at the expected
sampling rate. Replace the thimble
if you cannot and try again.
- Filter/thimble
holders should be preheated prior to the test run. Place them into the gas with the nozzle
pointed away from the flow for a minimum of 10 minutes prior to the start
of sampling.
- Preventing
Leaks
Duct Temps ≤ 400°F
·
Wrap the thimble
gasket with Teflon tape. This
helps seat thimble.
·
Tighten
stirrup by hand and snug with pliers. You want to make sure it is tight enough
so that thermal expansion does not cause a leak, but you also do not want to
crush the thimble by over-compressing the assembly.
Duct Temps > 400°F
·
Do not use Teflon tape
·
Remove
Teflon thrust ring inside Gelman ring.
·
Replace
Viton gasket in Gelman with a metal shim.
·
Replace regular thimbles with short thimble and spring loaded
stirrup system.
·
Consider placing the holders into the duct to
heat up, then remove and re-tighten prior to leak-checking.
·
Loosen holders immediately after final
leak-check to prevent seizing of the components when they cool.
- Vertical
sampling ports require the use of a “Shepard’s Hook” on the end of
the probe to orientate the thimble/filter holder upward. When assembling the sampling probe, orient the thimble/filter holder
between the Pitot tubes and Shepard’s Hook, and make sure that the pitots
or Shephard’s hook do not interfere with the nozzle.
- Anytime
you remove the probe from the duct, make sure you keep the nozzle opening
pointed slightly upward to
prevent loss of sample.
- After
the final leak-check and removal of the filter holder, elevate the
front-end of the probe and turn on the sample pump. This helps to drain any moisture
captured in the “dryer” and
probe. Walk-down the sampling line
from probe to the impingers or knock-outs prior to turning off the
sampling pump.