Method
301--Field Validation of Pollutant Measurement
Methods from
Various Waste Media - Appendix A - Test Methods
1.
APPLICABILITY AND PRINCIPLE
3.3 Surrogate Reference
Materials.
3.4 Isotopically
Labeled Materials.
4. EPA PERFORMANCE
AUDIT MATERIAL
5. PROCEDURE FOR
DETERMINATION OF BIAS AND PRECISION IN THE FIELD
5.2 Comparison Against
a Validated Test Method.
5.4 Probe Placement and
Arrangement For Stationary Source Stack or Duct Sampling.
6.2 Comparison with
Validated Method.
6.2.1 Paired Sampling
Systems.
6.2.2 Quadruplet
Replicate Sampling Systems.
6.3.3 Calculation of a
Correction Factor.
7. RUGGEDNESS TESTING
(OPTIONAL)
8. PROCEDURE FOR
INCLUDING SAMPLE STABILITY IN BIAS AND PRECISION EVALUATIONS
8.2.2 Other Waste Media
Testing.
9. PROCEDURE FOR
DETERMINATION OF PRACTICAL LIMIT OF QUANTITATION (OPTIONAL)
9.1 Practical Limit of
Quantitation.
9.2 Procedure I for
Estimating so.
9.3 Procedure II for
Estimating so.
10.0 FIELD VALIDATION
REPORT REQUIREMENTS
12. PROCEDURE FOR
OBTAINING A WAIVER
12.1.3
"Conditional" Test Methods.
This method, as
specified in the applicable subpart, is to be used whenever a source owner or
operator (hereafter referred to as an "analyst") proposes a test
method to meet a U.S. Environmental Protection Agency (EPA) requirement in the
absence of a validated method. This Method includes procedures for determining
and documenting the quality, i.e., systematic error (bias) and random error
(precision), of the measured concentrations from an effected source. This
method is applicable to various waste media (i.e., exhaust gas, wastewater,
sludge, etc.).
1.1.1 If EPA currently recognizes an appropriate
test method or considers the analyst's test method to be satisfactory for a
particular source, the Administrator may waive the use of this protocol or may
specify a less rigorous validation procedure. A list of validated methods may be obtained by contacting the Emission
Measurement Technical Information Center (EMTIC), Mail Drop 19, U.S.
Environmental Protection Agency,
Research Triangle Park, NC 27711, 919/541-0200. Procedures for obtaining
a waiver are in Section 12.0.
1.1.2 This method includes optional procedures
that may be used to expand the applicability of the proposed method. Section
7.0 involves ruggedness testing (Laboratory Evaluation), which demonstrates the
sensitivity of the method to various parameters. Section 8.0 involves a
procedure for including sample stability in bias and precision for assessing
sample recovery and analysis times; Section 9.0 involves a procedure for the
determination of the practical limit of quantitation for determining the lower
limit of the method. These optional procedures are required for the waiver
consideration outlined in Section 12.0.
The purpose of
these procedures is to determine bias and precision of a test method at the
level of the applicable standard. The procedures involve (a) introducing known
concentrations of an analyte or comparing the test method against a validated
test method to determine the method's bias and (b) collecting multiple or
collocated simultaneous samples to determine the method's precision.
Bias is
established by comparing the method's results against a reference value and may
be eliminated by employing a correction factor established from the data
obtained during the validation test. An offset bias may be handled accordingly.
Methods that have bias correction factors outside 0.7 to 1.3 are unacceptable.
Validated method to proposed method comparisons, Section 6.2, requires a more
restrictive test of central tendency and a lower correction factor allowance of
0.90 to 1.10.
At the minimum,
paired sampling systems shall be used to establish precision. The precision of
the method at the level of the standard shall not be greater than 50 percent
relative standard deviation. For a validated method to proposed method
equivalency comparisons, Section 6.2, the analyst must demonstrate that the
precision of the proposed test method is as precise as the validated method for
acceptance.
2.1 Negative bias. Bias Resulting when the
measured result is less than the "true" value.
2.2 Paired sampling system. A sampling system
capable of obtaining two replicate samples that were collected as closely as
possible in sampling time and sampling location.
2.3 Positive bias. Bias resulting when the
measured result is greater than the "true" value.
2.4 Proposed method. The sampling and
analytical methodology selected for field validation using the method described
herein.
2.5 Quadruplet sampling system. A sampling
system capable of obtaining four replicate samples that were collected as
closely as possible in sampling time and sampling location.
2.6 Surrogate compound. A compound that serves
as a model for the types of compounds being analyzed (i.e., similar chemical
structure, properties, behavior). The model can be distinguished by the method
from the compounds being analyzed.
The reference
materials shall be obtained or prepared at the level of the standard.
Additional runs with higher and lower reference material concentrations may be
made to expand the applicable range of the method, in accordance with the
ruggedness test procedures.
The analyst
shall obtain a known concentration of the reference material (i.e., analyte of
concern) from an independent source such as a specialty gas manufacturer,
specialty chemical company, or commercial laboratory. A list of vendors may be
obtained from EMTIC (see Section 1.1.1). The analyst should obtain the
manufacturer's stability data of the analyte concentration and recommendations
for recertification.
The analyst
shall obtain pure liquid components of the reference materials (i.e., analytes
of concern) from an independent manufacturer and dilute them in the same type
matrix as the source waste. The pure reference materials shall be certified by
the manufacturer as to purity and shelf life. The accuracy of all diluted
reference material concentrations shall be verified by comparing their response
to independently-prepared materials (independently prepared in this case means
prepared from pure components by a different analyst).
The analyst may
use surrogate compounds, e.g., for highly toxic or reactive organic compounds,
provided the analyst can demonstrate to the Administrator's satisfaction that
the surrogate compound behaves as the analyte. A surrogate may be an isotope or
one that contains a unique element (e.g., chlorine) that is not present in the
source or a derivative of the toxic or reactive compound, if the derivative
formation is part of the method's procedure. Laboratory experiments or
literature data may be used to show behavioral acceptability.
Isotope mixtures
may contain the isotope and the natural analyte. For best results, the isotope
labeled analyte concentration should be more than five times the natural
concentration of the analyte.
4.1 To assess the method bias independently,
the analyst shall use (in addition to the reference material) an EPA
performance audit material, if it is available. The analyst may contact EMTIC
(see Section 1.1.1) to receive a list of currently available EPA audit
materials. If the analyte is listed, the analyst should request the audit
material at least 30 days before the validation test. If an EPA audit material
is not available, request documentation from the validation report reviewing
authority that the audit material is currently not available from EPA. Include
this documentation with the field validation report.
4.2 The analyst shall sample and analyze the
performance audit sample three times according to the instructions provided
with the audit sample. The analyst shall submit the three results with the
field validation report. Although no acceptance criteria are set for these
performance audit results, the analyst and reviewing authority may use them to
assess the relative error of sample recovery, sample preparation, and
analytical procedures and then consider the relative error in evaluating the
measured emissions.
The analyst
shall select one of the sampling approaches below to determine the bias and
precision of the data. After analyzing the samples, the analyst shall calculate
the bias and precision according to the procedure described in Section 6.0.
When sampling a stationary source, follow the probe placement procedures in
Section 5.4.
This approach
shall be used only for methods that require mass spectrometry (MS) analysis.
Bias and precision are calculated by procedures described in Section 6.1.
5.1.1 Number
of Samples and Sampling Runs. Collect
a total of 12 replicate samples by either obtaining six sets of paired samples
or three sets of quadruplet samples.
5.1.2 Spiking
Procedure. Spike all 12
samples with the reference material at the level of the standard. Follow the
appropriate spiking procedures listed below for the applicable waste medium.
5.1.2.1
Exhaust Gas Testing. The
spike shall be introduced as close to the tip of the sampling probe as
possible.
5.1.2.1.1
Gaseous Reference Material with Sorbent or Impinger Sampling Trains. Sample the reference material (in the
laboratory or in the field) at a concentration which is close to the allowable
concentration standard for the time required by the method, and then sample the
gas stream for an equal amount of time. The time for sampling both the
reference material and gas stream should be equal; however, the time should be
adjusted to avoid sorbent breakthrough.
5.1.2.1.2
Gaseous Reference Material with Sample Container (Bag or Canister). Spike the sample containers after
completion of each test run with an amount equal to the allowable concentration
standard of the emission point. The final concentration of the reference
material shall approximate the level of the emission concentration in the
stack. The volume amount of reference material shall be less than 10 percent of
the sample volume.
5.1.2.1.3
Liquid and Solid Reference Material with Sorbent or Impinger Trains. Spike the trains with an amount equal to
the allowable concentration standard before sampling the stack gas. The spiking
should be done in the field; however, it may be done in the laboratory.
5.1.2.1.4
Liquid and Solid Reference Material with Sample Container (Bag or Canister). Spike the containers at the completion of
each test run with an amount equal to the level of the emission standard.
5.1.2.2 Other
Waste Media. Spike the 12
replicate samples with the reference material either before or directly after
sampling in the field.
Bias and
precision are calculated using the procedures described in Section 6.2. This
approach shall be used when a validated method is available and an alternative
method is being proposed.
5.2.1 Number
of Samples and Sampling Runs. Collect
nine sets of replicate samples using a paired sampling system (a total of 18
samples) or four sets of replicate samples using a quadruplet sampling system
(a total of 16 samples). In each sample set, the validated test method shall be
used to collect and analyze half of the samples.
5.2.2
Performance Audit Exception. Conduct
the performance audit as required in Section 4.0 for the validated test method.
Conducting a performance audit on the test method being evaluated is
recommended.
This approach
shall be used when Sections 5.1 and 5.2 are not applicable. Bias and precision
are calculated using the procedures described in Section 6.3.
5.3.1 Number
of Samples and Sampling Runs. Collect
a total of 24 samples using the quadruplet sampling system (a total of 6 sets
of replicate samples).
5.3.2 In each quadruplet set, spike half of the
samples (two out of the four) with the reference material according to the
applicable procedure in Section 5.1.2.1 or 5.1.2.2.
The probes shall
be placed in the same horizontal plane. For paired sample probes the
arrangement should be that the probe tip is 2.5 cm from the outside edge of the
other with a pitot tube on the outside of each probe. Other paired arrangements
for the pitot tube may be acceptable. For quadruplet sampling probes, the tips
should be in a 6.0 cm x 6.0 cm square area measured from the center line of the
opening of the probe tip with a single pitot tube in the center or two pitot
tubes with their location on either side of the probe tip configuration. An
alternative arrangement should be proposed when ever the cross-sectional area
of the probe tip configuration is approximately 5 percent of the stack or duct
cross-sectional area.
Data resulting
from the procedures specified in Section 5.0 shall be treated as follows to
determine bias, correction factors, relative standard deviations, precision, and
data acceptance.
Analyze the data
for isotopic spiking tests as outlined in Sections 6.1.1 through 6.1.6.
6.1.1 Calculate the numerical value of the bias
using the results from the analysis of the isotopically spiked field samples
and the calculated value of ENDFIELD the isotopically labeled spike:
where:
B = Bias at the
spike level.
Sm = Mean of the measured values of the
isotopically spiked samples.
CS = Calculated
value of the isotopically labeled spike.
6.1.2 Calculate the standard deviation of the Si values as follows:
where:
Si = Measured value of the isotopically
labeled analyte in the ith field sample,
n = Number of
isotopically spiked samples, 12.
6.1.3 Calculate the standard deviation of the
mean (SDM) as follows:
6.1.4 Test the bias for statistical significance
by calculating the t-statistic,
and compare it
with the critical value of the two-sided t-distribution at the 95-percent
confidence level and n-1 degrees of freedom. This critical value is 2.201 for
the eleven degrees of freedom when the procedure specified in Section 5.1.2 is
followed. If the calculated t-value is greater than the critical value the bias
is statistically significant and the analyst should proceed to evaluate the
correction factor.
6.1.5
Calculation of a Correction Factor. If
the t-test does not show that the bias is statistically significant, use all
analytical results without correction and proceed to the precision evaluation.
If the method's bias is statistically significant, calculate the correction
factor, CF, using the following equation:
If the CF is
outside the range of 0.70 to 1.30, the data and method are considered
unacceptable. For correction factors within the range, multiply all analytical
results by the CF to obtain the final values.
6.1.6
Calculation of the Relative Standard Deviation (Precision). Calculate the relative standard deviation
as follows:
where Sm is the measured mean of the isotopically
labeled spiked samples.
Analyze the data
for comparison with a validated method as outlined in Sections 6.2.1 or 6.2.2,
as appropriate. Conduct these procedures in order to determine if a proposed
method produces results equivalent to a validated method. Make all necessary
bias corrections for the validated method, as appropriate. If the proposed
method fails either test, the method results are unacceptable, and conclude
that the proposed method is not as precise or accurate as the validated method.
For highly variable sources, additional precision checks may be necessary. The
analyst should consult with the Administrator if a highly variable source is
suspected.
6.2.1.1
Precision. Determine the
acceptance of the proposed method's variance with respect to the variability of
the validated method results. If a significant difference is determined, the
proposed method and the results are rejected. Proposed methods demonstrating
F-values equal to or less than the critical value have acceptable precision.
6.2.1.2 Calculate the variance of the proposed
method, Sp 2 and the variance of the validated method, Sv 2, using the following equation:
where:
SDv = Standard deviation provided with the
validated method,
SDp = Standard deviation of the proposed method
calculated using Equation 301-9a.
6.2.1.3 The
F-test. Determine if the
variance of the proposed method is significantly different from that of the
validated method by calculating the F-value using the following equation:
Compare the
experimental F value with the critical value of F. The critical value is 1.0
when the procedure specified in section 5.2.1 for paired trains is followed. If
the calculated F is greater than the critical value, the difference in
precision is significant and the data and proposed method are unacceptable.
6.2.1.4 Bias
Analysis. Test the bias
for statistical significance by calculating the t-statistic and determine if
the mean of the differences between the proposed method and the validated
method is significant at the 80-percent confidence level. This procedure
requires the standard deviation of the validated method, SDv, to be known. Employ the value furnished
with the method. If the standard deviation of the validated method is not
available, the paired replicate sampling procedure may not be used. Determine
the mean of the paired sample
(If SDv > SDd,
let SD = SDd/1.414). Calculate the value of the
t-statistic using the following
equation:
where n is the
total number of paired samples. For the procedure in Section 5.2.1, n equals
nine. Compare the calculated t-statistic with the corresponding value from the
table of the t-statistic. When nine runs are conducted, as specified in Section
5.2.1, the critical value of the t-statistic is 1.397 for eight degrees of
freedom. If the calculated t-value is greater than the critical value the bias
is statistically significant and the analyst should proceed to evaluate the
correction factor.
6.2.1.5
Calculation of a Correction Factor. If
the statistical test cited above does not show a significant bias with respect
to the reference method, assume that the proposed method is unbiased and use
all analytical results without correction. If the method's bias is
statistically significant, calculate the correction factor, CF, as follows:
where Vm is the mean of the validated method's
values. Multiply all analytical results by CF to obtain the final values. The
method results, and the method, are unacceptable if the correction factor is
outside the range of 0.9 to 1.10.
6.2.2.1
Precision. Determine the
acceptance of the proposed method's variance with respect to the variability of
the validated method results. If a significant difference is determined the
proposed method and the results are rejected.
6.2.2.2 Calculate the variance of the proposed
method, Sp 2 using the following equation:
where the di's are the differences between the
validated method values and the proposed method values.
6.2.2.3 The
F-test. Determine if the
variance of the proposed method is more variable than that of the validated
method by calculating the F-value using Equation 301-8. Compare the
experimental F value with the critical value of F. The critical value is 1.0
when the procedure specified in section 5.2.2 for quadruplet trains is followed.
The calculated F should be less than or equal to the critical value. If the
difference in precision is significant the results and the proposed method are
unacceptable.
6.2.2.4 Bias
Analysis. Test the bias
for statistical significance at the 80 percent confidence level by calculating
the t-statistic. Determine the bias (mean of the differences between the
proposed method and the validated method, dm)
and the standard deviation, SDd, of the
differences. Calculate the standard deviation of the differences, SDd, using Equation 301-2 and substituting di for Si.
The following equation is used to calculate di:
and: V1i = First measured value of the validated
method in the ith test sample.
P1i = First measured value of the proposed
method in the ith test sample.
Calculate the
t-statistic using Equation 301-9 where n is the total number of test sample
differences (di). For the procedure in Section 5.2.2, n
equals four. Compare the calculated t-statistic with the corresponding value
from the table of the t-statistic and determine if the mean is significant at
the 80-percent confidence level. When four runs are conducted, as specified in
Section 5.2.2, the critical value of the t-statistic is 1.638 for three degrees
of freedom. If the calculated t-value is greater than the critical value the
bias is statistically significant and the analyst should proceed to evaluate
the correction factor.
6.2.2.5
Correction Factor Calculation. If
the method's bias is statistically significant, calculate the correction factor,
CF, using Equation 301-10. Multiply all analytical results by CF to obtain the
final values. The method results, and the method, are unacceptable if the
correction factor is outside the range of 0.9 to 1.10.
Analyze the data
for analyte spike testing as outlined in Sections 6.3.1 through 6.3.3.
6.3.1.1
Spiked Samples. Calculate
the difference, di, between the pairs of the spiked proposed
method measurements for each replicate sample set. Determine the standard deviation
(SDs) of the spiked values using the following
equation:
where: n =
Number of paired samples.
Calculate the
relative standard deviation of the proposed spiked method using Equation 301-6 where Sm is the measured mean of the analyte spiked
samples. The proposed method is unacceptable if the RSD is greater than 50
percent.
6.3.1.2
Unspiked Samples. Calculate
the standard deviation of the unspiked values using Equation 301-13 and the
relative standard deviation of the proposed unspiked method using Equation
301-6 where Sm
is the measured mean of the
unspiked samples. The RSD must be less than or equal to 50 percent.
Calculate the
numerical value of the bias using the results from the analysis of the spiked
field samples, the unspiked field samples, and the calculated value of the
spike:
where:
B = Bias at the
spike level.
Sm = Mean of the spiked samples.
Mm = Mean of the unspiked samples.
CS = Calculated
value of the spiked level.
6.3.2.1 Calculate the standard deviation of the mean
using the following equation where SDs and
SDu are the standard deviations of the spiked
and unspiked sample values respectively
6.3.2.2 Test the bias for statistical significance
by calculating the t- statistic using Equation 301-4 and comparing it with the
critical value of the two-sided t-distribution at the 95-percent confidence
level and n-1 degrees of freedom. This critical value is 2.201 for the eleven
degrees of freedom.
If the t-test
shows that the bias is not statistically significant, use all analytical
results without correction. If the method's bias is statistically significant,
calculate the correction factor using Equation 301-5. Multiply all analytical results
by CF to obtain the final values.
7.1.1 Ruggedness testing is a useful and
cost-effective laboratory study to determine the sensitivity of a method to
certain parameters such as sample collection rate, interferant concentration,
collecting medium temperature, or sample recovery temperature. This Section
generally discusses the principle of the ruggedness test. A more detailed
description is presented in citation 10 of Section 13.0.
7.1.2 In a ruggedness test, several variables are
changed simultaneously rather than one variable at a time. This reduces the
number of experiments required to evaluate the effect of a variable. For
example, the effect of seven variables can be determined in eight experiments
rather than 128 (W.J. Youden, Statistical
Manual of the Association of Official Analytical Chemists, Association of Official Analytical
Chemists, Washington, DC, 1975, pp. 33-36).
7.1.3 Data from ruggedness tests are helpful in
extending the applicability of a test method to different source concentrations
or source categories.
8.1.1 The test method being evaluated must
include procedures for sample storage and the time within which the collected
samples shall be analyzed.
8.1.2 This Section identifies the procedures for
including the effect of storage time in bias and precision evaluations. The
evaluation may be deleted if the test method specifies a time
for sample
storage.
The following
procedures shall be conducted to identify the effect of storage times on
analyte samples. Store the samples according to the procedure specified in the
test method. When using the analyte spiking procedures (Section 5.3), the study
should include equal numbers of spiked and unspiked samples.
8.2.1.1 For sample container (bag or canister) and
impinger sampling systems, Sections 5.1 and 5.3, analyze six of the samples at
the minimum storage time. Then analyze the same six samples at the maximum
storage time.
8.2.1.2 For sorbent and impinger sampling systems,
Sections 5.1 and 5.3, that require extraction or digestion, extract or digest
six of the samples at the minimum storage time and extract or digest six other
samples at the maximum storage time. Analyze an aliquot of the first six
extracts (digestates) at both the minimum and maximum storage times. This will
provide some freedom to analyze extract storage impacts.
8.2.1.3 For sorbent sampling systems, Sections 5.1
and 5.3, that require thermal desorption, analyze six samples at the minimum
storage time. Analyze another set of six samples at the maximum storage time.
8.2.1.4 For systems set up in accordance with
Section 5.2, the number of samples analyzed at the minimum and maximum storage
times shall be half those collected (8 or 9). The procedures for samples
requiring extraction or digestion should parallel those
in Section
8.2.1.
Analyze half of
the replicate samples at the minimum storage time and the other half at the
maximum storage time in order to identify the effect of storage times on
analyte samples.
9.1.1 The practical limit of quantitation (PLQ)
is the lowest level above which quantitative results may be obtained with an
acceptable degree of confidence. For this protocol, the PLQ is defined as 10
times the standard deviation, so, at the
blank level. This PLQ corresponds to an uncertainty of ±30 percent at the
99-percent confidence level.
9.1.2 The PLQ will be used to establish the lower
limit of the test method.
This procedure
is acceptable if the estimated PLQ is no more than twice the calculated PLQ. If
the PLQ is greater than twice the calculated PLQ use Procedure II.
9.2.1 Estimate the PLQ and prepare a test
standard at this level. The test standard could consist of a dilution of the
reference material described in Section 3.0.
9.2.2 Using the normal sampling and analytical
procedures for the method, sample and analyze this standard at least seven
times in the laboratory.
9.2.3 Calculate the standard deviation, so, of the measured values.
9.2.4 Calculate the PLQ as 10 times so.
This procedure
is to be used if the estimated PLQ is more than twice the calculated PLQ.
9.3.1 Prepare two additional standards at
concentration levels lower than the standard used in Procedure I.
9.3.2 Sample and analyze each of these standards
at least seven times.
9.3.3 Calculate the standard deviation for each
concentration level.
9.3.4 Plot the standard deviations of the three
test standards as a function of the standard concentrations.
9.3.5 Draw a best-fit straight line through the
data points and extrapolate to zero concentration. The standard deviation at
zero concentration is so.
9.3.6 Calculate the PLQ as 10 times so.
The field
validation report shall include a discussion of the regulatory objectives for
the testing which describe the reasons for the test, applicable emission
limits, and a description of the source. In addition, validation results shall
include:
10.1 Summary of the results and calculations
shown in Section 6.0.
10.2 Reference material certification and
value(s).
10.3 Performance audit results or letter from
the reviewing authority stating the audit material is currently not available.
10.4 Laboratory demonstration of the quality of
the spiking system.
10.5 Discussion of laboratory evaluations.
10.6 Discussion of field sampling.
10.7 Discussion of sample preparations and
analysis.
10.8 Storage times of samples (and extracts, if
applicable).
10.9 Reasons for eliminating any results.
The correction
factor calculated in Section 6.0 shall be used to adjust the sample
concentrations in all followup tests conducted at the same source. These tests
shall consist of at least three replicate samples, and the average shall be
used to determine the pollutant concentration. The number of samples to be
collected and analyzed shall be as follows, depending on the validated method
precision level:
11.1 Validated relative standard deviation (RSD)
< ±15 Percent. Three replicate samples.
11.2 Validated RSD < ±30 Percent. Six
replicate samples.
11.3 Validated RSD < ±50 Percent. Nine
replicate samples.
11.4 Equivalent method. Three replicate samples.
These procedures
may be waived or a less rigorous protocol may be granted for site-specific
applications. The following are three example situations for which a waiver may
be considered.
If the test
method has been validated previously at a "similar" source, the
procedures may be waived provided the requester can demonstrate to the
satisfaction of the Administrator that the sources are "similar." The
methods's applicability to the "similar" source may be demonstrated
by conducting a ruggedness test as described in Section 6.0.
In some cases,
bias and precision may have been documented through laboratory tests or
protocols different from this method. If the analyst can demonstrate to the
satisfaction of the Administrator that the bias and precision apply to a
particular application, the Administrator may waive these procedures or parts
of the procedures.
When the method
has been demonstrated to be valid at several sources, the analyst may seek a
"conditional" method designation from the Administrator.
"Conditional" method status provides an automatic waiver from the
procedures provided the test method is used within the stated applicability.
In general, the
requester shall provide a thorough description of the test method, the intended
application, and results of any validation or other supporting documents.
Because of the many potential situations in which the Administrator may grant a
waiver, it is neither possible nor desirable to prescribe the exact criteria
for a waiver. At a minimum, the requester is responsible for providing the
following.
12.2.1 A clearly written test method, preferably
in the format of 40 CFR 60, Appendix A Test Methods. The method must include an
applicability statement, concentration range, precision, bias (accuracy), and
time in which samples must be analyzed.
12.2.2.2 Summaries (see Section 10.0) of previous
validation tests or other supporting documents. If a different procedure from
that described in this method was used, the requester shall provide appropriate
documents substantiating (to the satisfaction of the Administrator) the bias
and precision values.
12.2.2.3 Results of testing conducted with respect
to Sections 7.0, 8.0, and 9.0.
12.2.3 Discussion of the applicability statement
and arguments for approval of the waiver. This discussion should address as
applicable the following: Applicable regulation, emission standards, effluent
characteristics, and process operations.
Each request
shall be in writing and signed by the analyst. Submit requests to the Director,
OAQPS, Technical Support Division, U.S. Environmental Protection Agency,
Research Triangle Park, NC 27711.
1. Albritton, J.R., G.B. Howe, S.B. Tompkins,
R.K.M. Jayanty, and C.E. Decker. 1989. Stability of Parts-Per-Million Organic
Cylinder Gases and Results of Source Test Analysis Audits, Status Report No.
11. Environmental Protection Agency Contract 68-02- 4125. Research Triangle
Institute, Research Triangle Park, NC. September.
2. DeWees, W.G., P.M. Grohse, K.K. Luk, and
F.E. Butler. 1989. Laboratory and Field Evaluation of a Methodology for
Speciating Nickel Emissions from Stationary Sources. EPA Contract 68-02-4442.
Prepared for Atmospheric Research and Environmental Assessment Laboratory,
Office of Research and Development, U.S. Environmental Protection Agency,
Research Triangle Park, NC 27711. January. 23
3. Keith, L.H., W. Crummer, J. Deegan Jr.,
R.A. Libby, J.K. Taylor, and G. Wentler. 1983. Principles of Environmental
Analysis. American Chemical Society, Washington, DC.
4. Maxwell, E.A. 1974. Estimating variances
from one or two measurements on each sample. Amer. Statistician 28:96-97.
5. Midgett, M.R. 1977. How EPA Validates NSPS
Methodology. Environ. Sci. & Technol. 11(7):655-659.
6. Mitchell, W.J., and M.R. Midgett. 1976.
Means to evaluate performance of stationary source test methods. Environ. Sci.
& Technol. 10:85-88.
7. Plackett, R.L., and J.P. Burman. 1946. The
design of optimum multifactorial experiments. Biometrika, 33:305.
8. Taylor, J.K. 1987. Quality Assurance of
Chemical Measurements. Lewis Publishers, Inc., pp. 79-81.
9. U.S. Environmental Protection Agency. 1978.
Quality Assurance Handbook for Air Pollution Measurement Systems: Volume III.
Stationary Source Specific Methods. Publication No. EPA-600/4-77-027b. Office
of Research and Development Publications, 26 West St. Clair St., Cincinnati, OH
45268.
10. U.S. Environmental Protection Agency. 1981.
A Procedure for Establishing Traceability of Gas Mixtures to Certain National
Bureau of Standards Standard Reference Materials. Publication No.
EPA-600/7-81-010. Available from the U.S. EPA, Quality Assurance Division
(MD-77), Research Triangle Park, NC 27711.
11. U.S. Environmental Protection Agency. 1991.
Protocol for The Field Validation of Emission Concentrations From Stationary
Sources. Publication No. 450/4-90-015. Available from the U.S. EPA, Emission
Measurement Technical Information Center, Technical Support Division (MD-14),
Research Triangle Park, NC 27711.
12. Youdon, W.J. Statistical techniques for
collaborative tests. In: Statistical Manual of the Association of Official
Analytical Chemists, Association of Official Analytical Chemists,
Washington, DC,
1975, pp. 33-36.