Method 23 -
Determination of Polychlorinated Dibenzo-p-dioxins and Polychlorinated
Dibenzofurans from Municipal Waste Combustors
1.
APPLICABILITY AND PRINCIPLE
3.1.6 Chromic Acid
Cleaning Solution.
5.2 Sample Cleanup and
Fractionation.
7.1 Sampling Train
Collection Efficiency Check.
7.2 Internal Standard
Percent Recoveries.
7.3 Surrogate Standard
Recoveries.
8.3 Audit Sample
Availability.
This method is
applicable to the determination of emissions of polychlorinated
dibenzo-p-dioxins (PCDD's) and polychlorinated dibenzofurans (PCDF's) from
stationary sources.
A sample is
withdrawn isokinetically from the gas stream and collected in the sample probe,
on a glass fiber filter, and on a packed column of adsorbent material. The
sample cannot be separated into a particle and vapor fraction. The PCDD's and
PCDF's are extracted from the sample, separated by high resolution gas chromatography
(HRGC), and measured by high resolution mass spectrometry (HRMS).
A schematic of the
sampling train is shown in Figure 23-1. Sealing greases may
not be used in assembling
the train. The train is identical to that described in Section 2.1 of Method 5 of this appendix with the following
additions:
2.1.1 Nozzle. The nozzle shall be made of nickel,
nickel-plated stainless steel, quartz, or borosilicate glass.
2.1.2 Sample
Transfer Lines. The sample
transfer lines, if needed, shall be heat traced, heavy walled TFE (1/2 in. OD
with 1/8 in. wall) with connecting fittings that are capable of forming leak
free, vacuum-tight connections without using sealing greases. The line shall be
as short as possible and must be maintained at 120¡C.
2.1.3 Filter
Support. Teflon or
Teflon-coated wire.
2.1.4
Condenser. Glass, coil type
with compatible fittings. A schematic diagram is shown in Figure
23-2.
2.1.5 Water
Bath. Thermostatically
controlled to maintain the gas temperature exiting the condenser at 20¡C
(68¡F).
2.1.6 Adsorbent
Module. Glass container to
hold the solid adsorbent. A schematic diagram is shown in Figure 23-2. Other
physical configurations of the resin trap/condenser assembly are acceptable.
The connecting fittings shall form leak-free, vacuum tight seals. No sealant
greases shall be used in the sampling train. A coarse glass frit is included to
retain the adsorbent.
2.2.1 Fitting
Caps. Ground glass, Teflon
tape, or aluminum foil (Section 2.2.6) to cap off the sample exposed sections
of the train and sorbent module.
Figure
23.2. Condenser and Absorbent Trap
2.2.2 Wash
Bottles. Teflon, 500-mL.
2.2.3 Probe Liner, Probe Nozzle, and Filter Holder
Brushes. Inert bristle brushes with pre-cleaned stainless steel or Teflon
handles. The probe brush shall have extensions of stainless steel or Teflon, at
least as long as the probe. The brushes shall be properly sized and shaped to
brush out the nozzle, probe liner, and transfer line, if used.
2.2.4 Filter
Storage Container. Sealed
filter holder, wide-mouth amber glass jar with Teflon-lined cap, glass Petri
dish.
2.2.5 Balance. Triple beam.
2.2.6 Aluminum
Foil. Heavy duty,
hexane-rinsed.
2.2.7 Metal
Storage Container. Air tight
container to store silica gel.
2.2.8 Graduated
Cylinder. Glass, 250-mL with
2-mL graduations.
2.2.9 Glass
Sample Storage Containers. Amber
glass bottles for sample glassware washes, 500- or 1000-mL, with leak free
Teflon-lined caps.
2.3.1 Sample
Containers. 125- and 250-mL
flint glass bottles with Teflon-lined caps.
2.3.2 Test
Tubes. Glass.
2.3.3 Soxhlet
Extraction Apparatus. Capable
of holding 43 x 123 mm extraction thimbles.
2.3.4
Extraction Thimble. Glass,
pre-cleaned cellulosic, or glass fiber.
2.3.5 Pasteur
Pipettes. For preparing
liquid chromatographic columns.
2.3.6
Reacti-vials. Amber glass, 2-mL,
silanized prior to use.
2.3.7 Rotary
Evaporator. Buchi/Brinkman
RF-121 or equivalent.
2.3.8
Nitrogen Evaporative Concentrator. N-Evap
Analytical Evaporator Model III or equivalent.
2.3.9
Separatory Funnels. Glass,
2-liter.
2.3.10 Gas Chromatograph.
Consisting of the following
components:
2.3.10.1 Oven. Capable of maintaining the separation column
at the proper operating temperature +10¡C and performing programmed
increases in temperature at rates of at least 40¡C/min.
2.3.10.2 Temperature
Gauges. To monitor column
oven, detector, and exhaust temperatures +1¡C.
2.3.10.3 Flow
Systems. Gas metering system
to measure sample, fuel, combustion gas, and carrier gas flows.
2.3.10.4
Capillary Columns. A fused
silica column, 60 x 0.25 mm inside diameter (ID), coated with DB-5 and a fused
silica column, 30 m x 0.25 mm ID coated with DB-225. Other column systems may
be substituted provided that the user is able to demonstrate, using calibration
and performance checks, that the column system is able to meet the
specifications of Section 6.1.2.2.
2.3.11 Mass
Spectrometer. Capable of
routine operation at a resolution of 1:10000 with a stability of +5 ppm.
2.3.12 Data
System. Compatible with the
mass spectrometer and capable of monitoring at least five groups of 25 ions.
2.3.13
Analytical Balance. To
measure within 0.1 mg.
Glass fiber
filters, without organic binder, exhibiting at least 99.95 percent efficiency
(<0.05 percent penetration) on 0.3-micron dioctyl phthalate smoke particles.
The filter efficiency test shall be conducted in accordance with ASTM Standard
Method D 2986-71 (Reapproved 1978) (incorporated by reference - see ¤60.17).
3.1.1.1
Pre-cleaning. All filters
shall be cleaned before their initial use. Place a glass extraction thimble and
1 g of silica gel and a plug of glass wool into a Soxhlet apparatus, charge the
apparatus with toluene, and reflux for a minimum of 3 hours. Remove the toluene
and discard it, but retain the silica gel. Place no more than 50 filters in the
thimble onto the silica gel bed and top with the cleaned glass wool. Charge the
Soxhlet with toluene and reflux for 16 hours. After extraction, allow the
Soxhlet to cool, remove the filters, and dry them under a clean nitrogen (N2) stream. Store the filters in a glass Petri
dish sealed with Teflon tape.
Amberlite XAD-2
resin. Thoroughly cleaned before initial use.
3.1.2.1
Cleaning. Procedure may be
carried out in a giant Soxhlet extractor. An all-glass filter thimble
containing an extra-coarse frit is used for extraction of XAD-2. The frit is
recessed 10-15 mm above a crenellated ring at the bottom of the thimble to
facilitate drainage. The resin must be carefully retained in the extractor cup
with a glass wool plug and a stainless steel ring because it floats on
methylene chloride. This process involves sequential extraction in the
following order.
3.1.2.2 Drying.
3.1.2.2.1
Drying Column. Pyrex pipe,
10.2 cm ID by 0.6 m long, with suitable retainers.
3.1.2.2.2
Procedure. The adsorbent
must be dried with clean inert gas. Liquid nitrogen from a standard commercial
liquid nitrogen cylinder has proven to be a reliable source for large volumes
of gas free from organic contaminants. Connect the liquid nitrogen cylinder to
the column by a length of cleaned copper tubing, 0.95 cm ID, coiled to pass
through a heat source. A convenient heat source is a water-bath heated from a
steam line. The final nitrogen temperature should only be warm to the touch and
not over 40¡C. Continue flowing nitrogen through the adsorbent until all the
residual solvent is removed. The flow rate should be sufficient to gently
agitate the particles, but not so excessive as to cause the particles to
fracture.
3.1.2.3 Quality
Control Check. The adsorbent
must be checked for residual toluene prior to use.
3.1.2.3.1
Extraction. Weigh a 1.0 g
sample of dried resin into a small vial, add 3 mL of toluene, cap the vial, and
shake it well.
3.1.2.3.2
Analysis. Inject a 2 ¥ l
sample of the extract into a gas chromatograph operated under the following
conditions:
Column: 6 ft
x 1/8 in stainless steel containing 10 percent OV-101ª on 100/120 Supelcoport.
Carrier Gas:
Helium at a rate of 30 mL/min.
Detector:
Flame ionization detector operated at a sensitivity of 4 x 10-11 A/mV.
Injection
Port Temperature: 250¡C.
Detector
Temperature: 305¡C.
Oven
Temperature: 30¡C for 4 min; programmed to rise at 40¡C/min until it reaches
250¡C; return to 30¡C after 17 minutes.
Compare the
results of the analysis to the results from the reference solution. Prepare the
reference solution by injecting 2.5 ¥ l
of methylene chloride into 100 mL of toluene. This corresponds to 100 ¥ g of methylene chloride per g of adsorbent.
The maximum acceptable concentration is 1000 ¥ g/g
of adsorbent. If the adsorbent exceeds this level, drying must be continued
until the excess methylene chloride is removed.
3.1.2.4
Storage. The adsorbent must
be used within 4 weeks of cleaning. After cleaning, the adsorbent may be stored
in a wide mouth amber glass container with a Teflon-lined cap or placed in
glass adsorbent modules tightly sealed with glass stoppers. If pre-cleaned
adsorbent is purchased in sealed containers, it must be used within 4 weeks
after the seal is broken.
Cleaned by
sequential immersion in three aliquots of methylene chloride, dried in a 110¡C
oven, and stored in a methylene chloride-washed glass container with a
Teflon-lined screw cap.
Deionized
distilled and stored in a methylene chloride-rinsed glass container with a
Teflon-lined screw cap.
Indicating type, 6
to 16 mesh. If previously used, dry at 175¡ C (350¡F) for two hours. New silica
gel may be used as received. Alternatively, other types of desiccants
(equivalent or better) may be used, subject to the approval of the
Administrator.
Dissolve 20 g of
sodium dichromate in 15 mL of water, and then carefully add 400 mL of concentrated
sulfuric acid.
3.2.1 Acetone. Pesticide quality.
3.2.2 Methylene
Chloride. Pesticide quality.
3.2.3 Toluene. Pesticide quality.
3.3.1 Potassium
Hydroxide. ACS grade,
2-percent (weight/volume) in water.
3.3.2 Sodium
Sulfate. Granulated, reagent
grade. Purify prior to use by rinsing with methylene chloride and oven drying.
Store the cleaned material in a glass container with a Teflon-lined screw cap.
3.3.3 Sulfuric
Acid. Reagent grade.
3.3.4 Sodium
Hydroxide. 1.0 N. Weigh 40 g
of sodium hydroxide into a 1-liter volumetric flask. Dilute to 1 liter with
water.
3.3.5 Hexane. Pesticide grade.
3.3.6 Methylene
Chloride. Pesticide grade.
3.3.7 Benzene. Pesticide grade.
3.3.8 Ethyl
Acetate.
3.3.9 Methanol.
Pesticide grade.
3.3.10 Toluene.
Pesticide grade.
3.3.11 Nonane. Pesticide grade.
3.3.12
Cyclohexane. Pesticide
Grade.
3.3.13 Basic
Alumina. Activity grade 1,
100-200 mesh. Prior to use, activate the alumina by heating for 16 hours at
130¡C. Store in a desiccator. Pre-activated alumina may be purchased from a
supplier and may be used as received.
3.3.14 Silica
Gel. Bio-Sil A, 100-200
mesh. Prior to use, activate the silica gel by heating for at least 30 minutes at
180¡C. After cooling, rinse the silica gel sequentially with methanol and
methylene chloride. Heat the rinsed silica gel at 50¡C for 10 minutes, then
increase the temperature gradually to 180¡C over 25 minutes and maintain it at
this temperature for 90 minutes. Cool at room temperature and store in a glass
container with a Teflon-lined screw cap.
3.3.15 Silica
Gel Impregnated with Sulfuric Acid. Combine
100 g of silica gel with 44 g of concentrated sulfuric acid in a screw capped
glass bottle and agitate thoroughly. Disperse the solids with a stirring rod
until a uniform mixture is obtained. Store the mixture in a glass container
with a Teflon-lined screw cap.
3.3.16 Silica
Gel Impregnated with Sodium Hydroxide. Combine 39 g of 1 N sodium hydroxide with 100 g of silica gel in
a screw capped glass bottle and agitate thoroughly. Disperse solids with a
stirring rod until a uniform mixture is obtained. Store the mixture in glass
container with a Teflon-lined screw cap.
3.3.17
Carbon/Celite. Combine 10.7
g of AX-21 carbon with 124 g of Celite 545 in a 250- mL glass bottle with a
Teflon-lined screw cap. Agitate the mixture thoroughly until a uniform mixture
is obtained. Store in the glass container.
3.3.18
Nitrogen. Ultra high purity.
3.3.19
Hydrogen. Ultra high purity.
3.3.20
Internal Standard Solution. Prepare
a stock standard solution containing the isotopic ally labeled PCDD's and
PCDF's at the concentrations shown in Table 1 under the heading "Internal
Standards" in 10 mL of nonane.
3.3.21
Surrogate Standard Solution. Prepare
a stock standard solution containing the isotopic ally labeled PCDD's and
PCDF's at the concentrations shown in Table 1 under the heading "Surrogate
Standards" in 10 mL of nonane.
3.3.22 Recovery
Standard Solution. Prepare a
stock standard solution containing the isotopic ally labeled PCDD's and PCDF's
at the concentrations shown in Table 1 under the heading "Recovery
Standards" in 10 mL of nonane.
The complexity of
this method is such that, in order to obtain reliable results, testers and
analysts should be trained and experienced with the procedures.
4.1.1 Pretest
Preparation.
4.1.1.1
Cleaning Glassware. All
glass components of the train upstream of and including the adsorbent module,
shall be cleaned as described in Section 3A of the "Manual of Analytical
Methods for the Analysis of Pesticides in Human and Environmental
Samples." Special care shall be devoted to the removal of residual
silicone grease sealants on ground glass connections of used glassware. Any
residue shall be removed by soaking the glassware for several hours in a
chromic acid cleaning solution prior to cleaning as described above.
4.1.1.2
Adsorbent Trap. The traps
must be loaded in a clean area to avoid contamination. They may not be loaded
in the field. Fill a trap with 20 to 40 g of XAD-2. Follow the XAD-2 with glass
wool and tightly cap both ends of the trap. Add 100 ¥ l of the surrogate standard solution (Section
3.3.21) to each trap.
4.1.1.3
Sampling Train. It is suggested
that all components be maintained according to the procedure described in
APTD-0576.
4.1.1.4 Silica
Gel. Weigh several 200 to
300 g portions of silica gel in airtight containers to the nearest 0.5 g.
Record the total weight of the silica gel plus container, on each container. As
an alternative, the silica gel may be weighed directly in its impinger or
sampling holder just prior to sampling.
4.1.1.5 Filter.
Check each filter against
light for irregularities and flaws or pinhole leaks. Pack the filters flat in a
clean glass container.
4.1.2
Preliminary Determinations. Same
as Section 4.1.2 of Method 5.
4.1.3
Preparation of Collection Train.
4.1.3.1 During preparation and assembly of the
sampling train, keep all train openings where contamination can enter, sealed
until sampling is about to begin.
4.1.3.2 Place approximately 100 mL of water in the
second and third impingers, leave the first and fourth impingers empty, and
transfer approximately 200 to 300 g of pre-weighed silica gel from its container
to the fifth impinger.
4.1.3.3 Place the silica gel container in a clean
place for later use in the sample recovery. Alternatively, the weight of the
silica gel plus the fifth impinger may be determined to the nearest 0.5 g and
recorded.
4.1.3.4 Assemble the sampling train as shown in Figure 23-1.
4.1.3.5 Turn on the adsorbent module and condenser
coil recirculating pump and begin monitoring the adsorbent module gas entry
temperature. Ensure proper sorbent gas entry temperature before proceeding and
before sampling is initiated. It is extremely important that the XAD-2
adsorbent resin temperature never exceed 50¡C because thermal decomposition
will occur. During testing, the XAD-2 temperature must not exceed 20¡C for efficient
capture of the PCDD's and PCDF's.
4.1.4
Leak-Check Procedure. Same
as Method 5, Section 4.1.4.
4.1.5 Sampling
Train Operation. Same as
Method 5, Section 4.1.5.
Proper cleanup
procedure begins as soon as the probe is removed from the stack at the end of
the sampling period. Seal the nozzle end of the sampling probe with Teflon tape
or aluminum foil. When the probe can be safely handled, wipe off all external
particulate matter near the tip of the probe. Remove the probe from the train
and close off both ends with aluminum foil. Seal off the inlet to the train
with Teflon tape, a ground glass cap, or aluminum foil. Transfer the probe and
impinger assembly to the cleanup area. This area shall be clean and enclosed so
that the chances of losing or contaminating the sample are minimized. Smoking,
which could contaminate the sample, shall not be allowed in the cleanup area.
Inspect the train prior to and during disassembly and note any abnormal
conditions, e.g., broken filters, colored impinger liquid, etc. Treat the
samples as follows:
4.2.1 Container
No. 1. Either seal the
filter holder or carefully remove the filter from the filter holder and place
it in its identified container. Do not place the filter in aluminum foil. Use a
pair of cleaned tweezers to handle the filter. If it is necessary to fold the
filter, do so such that the particulate cake is inside the fold. Carefully
transfer to the container any particulate matter and filter fibers which adhere
to the filter holder gasket, by using a dry inert bristle brush and a
sharp-edged blade. Seal the container.
4.2.2 Adsorbent
Module. Remove the module
from the train, tightly cap both ends, label it, and store it on ice for
transport to the laboratory.
4.2.3 Container
No. 2. Quantitatively
recover material deposited in the nozzle, probe transfer lines, the front half
of the filter holder, and the cyclone, if used, first, by brushing while
rinsing three times with acetone and then, by rinsing the probe three times
with methylene chloride. Collect all the rinses in Container No. 2. Rinse the
back half of the filter holder three times with acetone. Rinse the connecting
line between the filter and the condenser three times with acetone. Soak the
connecting line with three separate portions of methylene chloride for 5
minutes each. If using a separate condenser and adsorbent trap, rinse the
condenser in the same manner as the connecting line. Collect all the rinses in
Container No. 2 and mark the level of the liquid on the container.
4.2.4 Container
No. 3. Repeat the methylene
chloride-rinsing described in Section 4.2.3 using toluene as the rinse solvent.
Collect the rinses in Container No. 3 and mark the level of the liquid on the
container.
4.2.5 Impinger
Water. Measure the liquid in
the first four impingers to within 1 mL by using a graduated cylinder or by
weighing it to within 0.5 g by using a balance. Record the volume or weight of
liquid present. This information is required to calculate the moisture content
of the effluent gas. Discard the liquid after measuring and recording the
volume or weight.
4.2.7 Silica
Gel. Note the color of the
indicating silica gel to determine if it has been completely spent and make a mention
of its condition. Transfer the silica gel from the fifth impinger to its
original container and seal.
All glassware
shall be cleaned as described in Section 3A of the "Manual of Analytical
Methods for the Analysis of Pesticides in Human and Environmental
Samples." All samples must be extracted within 30 days of collection and
analyzed within 45 days of extraction.
5.1.1
Extraction System. Place an
extraction thimble (Section 2.3.4), 1 g of silica
gel, and a plug of glass wool into the Soxhlet apparatus, charge the apparatus
with toluene, and reflux for a minimum of 3 hours. Remove the toluene and
discard it, but retain the silica gel. Remove the extraction thimble from the
extraction system and place it in a glass beaker to catch the solvent rinses.
5.1.2 Container
No. 1 (Filter). Transfer the
contents directly to the glass thimble of the extraction system and extract
them simultaneously with the XAD-2 resin.
5.1.3 Adsorbent
Cartridge. Suspend the
adsorbent module directly over the extraction thimble in the beaker (See
Section 5.1.1). The glass frit of the module should be in the up position.
Using a Teflon squeeze bottle containing toluene, flush the XAD-2 into the
thimble onto the bed of cleaned silica gel. Thoroughly rinse the glass module
catching the rinsings in the beaker containing the thimble. If the resin is
wet, effective extraction can be accomplished by loosely packing the resin in
the thimble. Add the XAD-2 glass wool plug to the thimble.
5.1.4 Container
No. 2 (Acetone and Methylene Chloride). Concentrate the sample to a volume of about 1-2 mL using the
rotary evaporator apparatus at a temperature of less than 37¡C. Rinse the
sample container three times with small portions of methylene chloride and add
these to the concentrated solution and concentrate further to near dryness.
This residue contains particulate matter removed in the rinse of the sampling
train probe and nozzle. Add the concentrate to the filter and the AD-2 resin in
the Soxhlet apparatus described in Section 5.1.1.
5.1.5
Extraction. Add100 ¥ l of the internal standard solution (Section 3.3.20) to the extraction thimble containing
the contents of the adsorbent cartridge, the contents of Container No. 1, and
the concentrate from Section 5.1.4. Cover the contents of the extraction
thimble with the cleaned glass wool plug to prevent the XAD-2 resin from
floating into the solvent reservoir of the extractor. Place the thimble in the
extractor, and add the toluene contained in the beaker to the solvent
reservoir. Pour additional toluene to fill the reservoir approximately 2/3
full. Add Teflon boiling chips and assemble the apparatus. Adjust the heat
source to cause the extractor to cycle three times per hour. Extract the sample
for 16 hours. After extraction, allow the Soxhlet to cool. Transfer the toluene
extract and three 10-mL rinses to the rotary evaporator. Concentrate the
extract to approximately 10 mL. At this point the analyst may choose to split
the sample in half. If so, split the sample, store one half for future use, and
analyze the other half according to the procedures in Sections 5.2 and 5.3. In
either case, use a nitrogen evaporative concentrator to reduce the volume of
the sample being analyzed to near dryness. Dissolve the residue in 5 mL of
hexane.
5.1.6 Container
No. 3 (Toluene Rinse). Add
100 ¥ l of the internal standard solution (Section
3.3.20) to the contents of the container. Concentrate the sample to a volume of
about 1-5 mL using the rotary evaporator apparatus at a temperature of less
than 37¡C. Rinse the sample container three times with small portions of
toluene and add these to the concentrated solution and concentrate further to
near dryness. Analyze the extract separately according to the procedures in
Sections 5.2 and 5.3, but concentrate the solution in a rotary evaporator
apparatus rather than a nitrogen evaporative concentrator.
5.2.1 Silica Gel
Column. Pack one end of a
glass column, 20 mm x 230 mm, with glass wool. Add in sequence, 1 g silica gel,
2 g of sodium hydroxide impregnated silica gel, 1 g silica gel, 4 g of
acid-modified silica gel, and 1 g of silica gel. Wash the column with 30 mL of
hexane and discard. Add the sample extract, dissolved in 5 mL of hexane to the
column with two additional 5-mL rinses. Elute the column with an additional 90
mL of hexane and retain the entire eluate. Concentrate this solution to a
volume of about 1 mL using the nitrogen evaporative concentrator (Section 2.3.8).
5.2.2 Basic
Alumina Column. Shorten a
25-mL disposable Pasteur pipette to about 16 mL. Pack the lower section with
glass wool and 12 g of basic alumina. Transfer the concentrated extract from
the silica gel column to the top of the basic alumina column and elute the
column sequentially with 120 mL of 0.5 percent methylene chloride in hexane
followed by 120 mL of 35 percent methylene chloride in hexane. Discard the
first 120 mL of eluate. Collect the second 120 mL of eluate and concentrate it
to about 0.5 mL using the nitrogen evaporative concentrator.
5.2.3 AX-21
Carbon/Celite 545 Column. Remove
the bottom 0.5 in. from the tip of a 9-mL disposable Pasteur pipette. Insert a
glass fiber filter disk or glass wool plug in the top of the pipette 2.5 cm
from the constriction. Add sufficient carbon/Celiteª mixture to form a 2 cm
column (the 0.6 mL mark column. Top with a glass wool plug. In some cases AX-21
carbon fines may wash through the glass wool plug and enter the sample. This
may be prevented by adding a celite plug to the exit end of the column. Rinse
the column in sequence with 2 mL of 50 percent benzene in ethyl acetate, 1 mL
of a 50 percent methylene chloride in cyclohexane mixture, and 2 mL of hexane.
Discard these rinses. Transfer, the concentrate in 1 mL hexane from the basic
alumina column to the carbon/celite along with 1 ml of hexane rinse. Elute the
column sequentially with 2 mL of 50 percent methylene chloride in hexane and 2
mL of 50 percent benzene in ethyl acetate and discard the eluates. Invert the
column and elute in the reverse direction with 13 mL of toluene. Collect this
eluate. Concentrate the eluate in a rotary evaporator at 50¡C to about 1 mL.
Transfer the concentrate to a Reacti-vial using a toluene rinse and concentrate
to a volume of 200 µl using a stream of N2.
Store extracts at room temperature, shielded from light, until the analysis is
performed.
Analyze the sample
with a gas chromatograph coupled to a mass spectrometer (GC/MS) using the
instrumental parameters in Sections 5.3.1 and 5.3.2. Immediately prior to
analysis, add a 20 ¥
l aliquot of the recovery
standard solution from Table 1 to each sample. A 2 ¥ l aliquot of the extract is injected into the
GC. Sample extracts are first analyzed using the DB-5 capillary column to
determine the concentration of each isomer of PCDD's and PCDF's (tetra-through
octa-). If tetrachlorinated dibenzofurans are detected in this analysis, then
analyze another aliquot of the sample in a separate run, using the DB-225
column to measure the 2,3,7,8 tetra-chloro dibenzofuran isomer. Other column
systems may be used, provided that the user is able to demonstrate using
calibration and performance checks that the column system is able to meet the
specifications of Section 6.1.2.2.
5.3.1 Gas
Chromatograph Operating Conditions.
5.3.1.1
Injector. Configured for
capillary column, splitless, 250 ¡C.
5.3.1.2 Carrier
Gas. Helium, 1-2 ml/min.
5.3.1.3 Oven. Initially at 150 ¡C. Raise by at least 40
¡C/min to 190 ¡C and then by ¡C/min up to 300 ¡C.
5.3.2 High
Resolution Mass Spectrometer.
5.3.2.1
Resolution. 10,000 m/e.
5.3.2.2
Ionization Mode. Electron
impact.
5.3.2.3 Source
Temperature 250¡C.
5.3.2.4
Monitoring Mode. Selected
ion monitoring. A list of the various ions to be monitored is presented in
Table 3.
5.3.2.5
Identification Criteria. The
following identification criteria shall be used for the characterization of
polychlorinated dibenzodioxins and dibenzofurans.
1. The integrated
ion-abundance ratio (M/M+2 or M+2/M+4) shall be within 15 percent of the
theoretical value. The acceptable ion-abundance ratio ranges (+15%) for
the identification of chlorine-containing compounds are given in Table 4.
2. The retention
time for the analytes must be within 3 seconds of the corresponding 13C-labeled internal standard or surrogate
standard.
3. The monitored
ions, shown in Table 3 for a given analyte, shall reach
their maximum within 2 seconds of each other.
4. The
identification of specific isomers that do not have corresponding 13C-labeled standards is done by comparison of
the relative retention time (RRT) of the analyte to the nearest internal
standard retention time with reference (i.e., within 0.005 RRT units) to the
comparable RRT's found in the continuing calibration.
5. The signal to
noise ratio for all monitored ions must be greater than 2.5.
6. The
confirmation of 2, 3, 7, 8-TCDF shall satisfy all of the above identification
criteria.
7. For the
identification of PCDFÕs, no signal may be found in the corresponding PCDPE channels.
5.3.2.6
Quantification. The peak
areas for the two ions monitored for each analyte are summed to yield the total
response for each analyte. Each internal standard is used to quantify the
indigenous PCDD's or PCDF's in its homologous series. For example, the 13C12-2,3,7,8-tetra
chlorinated dibenzodioxin is used to calculate the concentrations of all other
tetra chlorinated isomers. Recoveries of the tetra- and penta- internal
standards are calculated using the 13C12-1,2,3,4-TCDD. Recoveries of the hexa-
through octa- internal standards are calculated using 13C12-1,2,3,7,8,9-HxCDD.
Recoveries of the surrogate standards are calculated using the corresponding
homolog from the internal standard.
Same as Method 5
with the following additions.
6.1.1 Initial
Calibration. Calibrate the
GC/MS system using the set of five standards shown in Table
2. The relative standard deviation for the mean response factor from each
of the unlabeled analytes (Table 2) and of the internal and surrogate standards
shall be less than or equal to the values in Table 5.
The signal to noise ratio for the GC signal present in every selected ion
current profile shall be greater than or equal to 2.5. The ion abundance ratios
shall be within the control limits in Table 4.
6.1.2 Daily
Performance Check.
6.1.2.1
Calibration Check. Inject
one ¥ l of solution Number 3 from Table 2. Calculate
the relative response factor (RRF) for each compound and compare each RRF to
the corresponding mean RRF obtained during the initial calibration. The
analyzer performance is acceptable if the measured RRF's for the labeled and
unlabeled compounds for the daily run are within the limits of the mean values
shown in Table 5. In addition, the ion-abundance ratios shall be within the
allowable control limits shown in Table 4.
6.1.2.2
Column Separation Check. Inject
a solution of a mixture of PCDD's and PCDF's that documents resolution between
2,3,7,8-TCDD and other TCDD isomers. Resolution is defined as a valley between
peaks that is less than 25 percent of the lower of the two peaks. Identify and
record the retention time windows for each homologous series. Perform a similar
resolution check on the confirmation column to document the resolution between
2,3,7,8 TCDF and other TCDF isomers.
Set mass
spectrometer lock channels as specified in Table 3. Monitor the quality control
check channels specified in Table 3 to verify instrument stability during the
analysis.
Add 100 ¥ l of the surrogate standards in Table 1 to
the adsorbent cartridge of each train before collecting the field samples.
A group of nine
carbon-labeled PCDDs and PCDFs representing the tetra- through octachlorinated
homologues, is added to every sample prior to extraction. The role of the
internal standards is to quantify the native PCDD's and PCDF's present in the
sample as well as to determine the overall method efficiency. Recoveries of the
internal standards must be between 40 to 130 percent for the tetra- through
hexachlorinated compounds while the range is 25 to 130 percent for the hepta-
and octachlorinated homologues.
The five surrogate
compounds in Table 2 are added to the resin in the adsorbent sampling cartridge
before the sample is collected. The surrogate recoveries are measured relative
to the internal standards and are a measure of the collection efficiency. They
are not used to measure the native PCDD's and PCDF's. All recoveries shall be
between 70 and 130 percent. Poor recoveries for all the surrogates may be an
indication of breakthrough in the sampling train. If the recovery of all
standards is below 70 percent, the sampling runs must be repeated. As an
alternative, the sampling runs do not have to be repeated if the final results
are divided by the fraction of surrogate recovery. Poor recoveries of isolated
surrogate compounds should not be grounds for rejecting an entire set of
samples.
Report the results
of the toluene QA rinse separately from the total sample catch. Do not add it
to the total sample.
When the method is
used to analyze samples to demonstrate compliance with a source emission
regulation, an audit sample must be analyzed, subject to availability.
Analyze an audit
sample with each set of compliance samples. The audit sample contains tetra
through octa isomers of PCDD and PCDF. Concurrently analyze the audit sample
and a set of compliance samples in the same manner to evaluate the technique of
the analyst and the standards preparation. The same analyst, analytical
reagents, and analytical system shall be used both for the compliance samples
and the EPA audit sample.
Audit samples will
be supplied only to enforcement agencies for compliance tests. Audit samples
may be obtained by writing:
Source Test
Audit Coordinator (MD-77B)
Quality
Assurance Division
Atmospheric
Research and Exposure Assessment Laboratory
U.S.
Environmental Protection Agency
Research
Triangle Park, NC 27711
or by calling the
Source Test Audit Coordinator (STAC) at (919) 541-7834. The audit sample
request must be made at least 30 days prior to the scheduled compliance sample
analysis.
Calculate the
audit sample concentration according to the calculation procedure provided in
the audit instructions included with the audit sample. Fill in the audit sample
concentration and the analyst's name on the audit response form included with
the audit instructions. Send one copy to the EPA Regional Office or the
appropriate enforcement agency and a second copy to the STAC. The EPA Regional
office or the appropriate enforcement agency will report the results of the
audit to the laboratory being audited. Include this response with the results
of the compliance samples in relevant reports to the EPA Regional Office or the
appropriate enforcement agency.
Same as Method 5, Section 6 with the following
additions.
9.1
Nomenclature.
9.2 Average
Relative Response Factor.
9.3
Concentration of the PCDD's and PCDF's.
9.4 Recovery
Standard Response Factor.
9.5 Recovery of
Internal Standards (R*).
9.6 Surrogate
Compound Response Factor.
9.7 Recovery of
Surrogate Compounds (Rs).
9.8 Minimum
Detectable Limit (DL).
9.9 Total
Concentration of PCDD's and PCDF's in the Sample.
Any PCDDs or PCDFs
that are reported as nondetected (below the DL) shall be counted as zero for
the purpose of calculating the total concentration of PCDDs and PCDFs in the
sample.
1. American
Society of Mechanical Engineers. Sampling for the Determination of Chlorinated
Organic Compounds in Stack Emissions. Prepared for U.S. Department of Energy
and U.S. Environmental Protection Agency. Washington DC. December 1984. 25 p.
2. American
Society of Mechanical Engineers. Analytical Procedures to Assay Stack Effluent
Samples and Residual Combustion Products for Polychlorinated Dibenzo-p-Dioxins
(PCDD) and Polychlorinated Dibenzofurans (PCDF). Prepared for the U.S.
Department of Energy and U.S. Environmental Protection Agency. Washington, DC.
December 1984. 23 p.
3. Thompson, J. R.
(ed.). Analysis of Pesticide Residues in Human and Environmental Samples. U.S. Environmental
Protection Agency. Research Triangle Park, NC. 1974.
4. Triangle
Laboratories. Case Study: Analysis of Samples for the Presence of Tetra Through
Octachloro-p-Dibenzodioxins and Dibenzofurans. Research Triangle Park, NC.
1988. 26 p.
5. U.S. Environmental
Protection Agency. Method 8290 - The Analysis of Polychlorinated
Dibenzo-p-dioxin and Polychlorinated Dibenzofurans by High-Resolution Gas
Chromatography/High-Resolution Mass Spectrometry. In: Test Methods for
Evaluating Solid Waste. Washington, DC. SW-846.
TABLE
23-1. COMPOSITION OF THE SAMPLE FORTIFICATION AND RECOVERY STANDARDS SOLUTIONS
TABLE
23-2. COMPOSITION OF THE INITIAL CALIBRATION SOLUTIONS
TABLE 23-3. ELEMENTAL COMPOSITIONS AND EXACT MASSES OF THE IONS MONITORED BY HIGH RESOLUTION MASS SPECTROMETRY FOR PCDD's AND PCDF's
TABLE 23-4. ACCEPTABLE RANGES FOR ION-ABUNDANCE RATIOS OF PCDD's AND PCDF's
TABLE 23-5. MINIMUM REQUIREMENTS FOR INITIAL AND DAILY CALIBRATION RESPONSE FACTORS