Technician Guidance Document
Setting up Leak Check Parameters
DelayLeakCheck : Is the time that unit
waits after it hits the target vacuum before it starts the actual
leak rate count. The time is in seconds and can be set via PDA
->Maintenance->System Settings. The max time is 3
minutes, but essentially you can use larger values if you edit the
INI file directly.
Delay Step Increases: Is the delay in
milliseconds between each increase in leak valve position. The
goal is to wait long enough to allow the vacuum to settle before
continuing to change the valve position. 200ms has worked in the past.
A change in this time may also require a change in LeakValveStep.
LeakValveStep: Percent increase in leak valve position after each DelayStepIncrease.
Example for Short HSL
Example for Long HSL